Comparison of central corneal thickness measurements with a new optical device and a standard ultrasonic pachymeter.

Nemeth G, Tsorbatzoglou A, Kertesz K, Vajas A, Berta A, Módis L

Department of Ophthalmology, Medical and Health Science Center, University of Debrecen, Debrecen, Hungary. nemeth222@yahoo.com

PURPOSE: To compare central corneal thickness (CCT) values obtained with ultrasonic pachymetry and a new optical method using partial coherence interferometry (PCI). SETTING: Department of Ophthalmology, Medical Health and Science Center, University of Debrecen, Debrecen, Hungary. METHODS: The study comprised 136 eyes of 70 patients whose spherical refractive error was not greater than +/-6.0 diopters (D) and whose keratometric astigmatism was not greater than 2.0 D. Central corneal thickness was measured 5 times with a new optical device (ACMaster, Zeiss) and with an ultrasonic pachymeter (AL-2000, Tomey). All measurements were obtained by the same investigator. RESULTS: Mean CCT was 531.2 microm +/- 3.9 (SD) with PCI and 547.8 +/- 36.0 microm with the ultrasonic device. The difference between groups was significant (P = .001). There was no difference between CCT values measured in right and left eyes (P = .55) with ultrasonography and PCI (P = .67). The coefficient variation was 0.73% for PCI and 6.5% for ultrasonography. Correlation between the CCT measurements with both devices was strong and statistically significant (Spearman correlation = .91, P = .001). CONCLUSIONS: Mean CCT values measured by the PCI method were significantly smaller than those measured by the ultrasonic device. Central corneal thickness measured with PCI is more reproducible and seems to be more reliable than that measured by ultrasonography.

Published 24 April 2006 in J Cataract Refract Surg, 32(3): 460-3.
Full-text of this article is available online (may require subscription).

Place a permanent text-link or advertisement here for just US$15.

© 2005-2008 Astigmatism Research Today. All Rights Reserved.



Astigmatism Research Today Archive:

Volume 1 (2005)
  Issue 1 (November)
  Issue 2 (December)

Volume 2 (2006)
  Issue 1 (January)
  Issue 2 (February)
  Issue 3 (March)
  Issue 4 (April)
  Issue 5 (May)
  Issue 6 (June)
  Issue 7 (July)
  Issue 8 (August)
  Issue 9 (September)
  Issue 10 (October)
  Issue 11 (November)
  Issue 12 (December)

Volume 3 (2007)
  Issue 1 (January)
  Issue 2 (February)
  Issue 3 (March)
  Issue 4 (April)
  Issue 5 (May)
  Issue 6 (June)
  Issue 7 (July)
  Issue 8 (August)
  Issue 9 (September)
  Issue 10 (October)
  Issue 11 (November)
  Issue 12 (December)

Volume 4 (2008)
  Issue 1 (January)
  Issue 2 (February)
  Issue 3 (March)
  Issue 4 (April)
  Issue 5 (May)
  Issue 6 (June)
  Issue 7 (July)
  Issue 8 (August)
  Issue 9 (September)
  Issue 10 (October)



Astigmatism Books

Refractive Eye Surgery<br>A Consumer's Complete Guide: LASIK, IntraLASIK, Epi-LASIK, CK, Implantable Contact Lenses, and Other Surgical Eye Procedures ... Dependence on Glasses and Contact Lenses

Refractive Eye Surgery
A Consumer's Complete Guide: LASIK, IntraLASIK, Epi-LASIK, CK, Implantable Contact Lenses, and Other Surgical Eye Procedures ... Dependence on Glasses and Contact Lenses